Detector 4 anneal 2012

From RHESSI Wiki

Revision as of 22:37, 12 December 2012 by NicoleDuncan (Talk | contribs)
Jump to: navigation, search

D4 has been spontaneously segmenting and desegmenting for approximately the past 30 days, it began on day 317. The high voltage was raised by 5 steps to stabilize the segmentation, and it appears to be initially successful. Brian Dennis made some plots to show that its resolution is as good as before the anneal. He also showed that the calibration used in SSW is not up to date.

G4 Front Slow 0x70 to 0x0C.

G4 Front Fast 0x60 to 0x50.