Detector 7 anneal 2012

From RHESSI Wiki

(Difference between revisions)
Jump to: navigation, search
(Added tables of threshold changes (none))
Line 1: Line 1:
 +
===Notes===
 +
There was an arc that led to a long period of extended noise in D7.  See [[Tohban_Report_2013-07-24]] for details.
There was an arc that led to a long period of extended noise in D7.  See [[Tohban_Report_2013-07-24]] for details.
   D7 HV was stepped down by 10 steps because of this.  (HV adjustment was on July 24, 2013 (day 205) at 19:29:15 UTC).  HV after the stepdown was ~3800 V.  This made no change in the noise level.
   D7 HV was stepped down by 10 steps because of this.  (HV adjustment was on July 24, 2013 (day 205) at 19:29:15 UTC).  HV after the stepdown was ~3800 V.  This made no change in the noise level.
   Update as of Aug. 7 2013: David has determined that the noise is not affecting D7 spectroscopy (although the calibration may have changed slightly), and livetime is not being significantly affected, so no further steps will be taken for now.
   Update as of Aug. 7 2013: David has determined that the noise is not affecting D7 spectroscopy (although the calibration may have changed slightly), and livetime is not being significantly affected, so no further steps will be taken for now.
 +
 +
===D7 front slow LLD threshold history===
 +
{| class="wikitable"
 +
|-
 +
| post 2012 anneal || 0x0C
 +
|}
 +
 +
===D7 front fast LLD threshold history===
 +
{| class="wikitable"
 +
|-
 +
| post 2012 anneal || 0x22
 +
|}
 +
 +
===D7 rear slow LLD threshold history===
 +
{| class="wikitable"
 +
|-
 +
| post 2012 anneal || 0x30
 +
|}
 +
 +
===D7 rear fast LLD threshold history===
 +
{| class="wikitable"
 +
|-
 +
| post 2012 anneal || 0x45
 +
|}

Revision as of 23:36, 17 March 2014

Contents

Notes

There was an arc that led to a long period of extended noise in D7. See Tohban_Report_2013-07-24 for details.

 D7 HV was stepped down by 10 steps because of this.  (HV adjustment was on July 24, 2013 (day 205) at 19:29:15 UTC).  HV after the stepdown was ~3800 V.  This made no change in the noise level.
 Update as of Aug. 7 2013: David has determined that the noise is not affecting D7 spectroscopy (although the calibration may have changed slightly), and livetime is not being significantly affected, so no further steps will be taken for now.

D7 front slow LLD threshold history

post 2012 anneal 0x0C

D7 front fast LLD threshold history

post 2012 anneal 0x22

D7 rear slow LLD threshold history

post 2012 anneal 0x30

D7 rear fast LLD threshold history

post 2012 anneal 0x45