* 2012 December 19: The noise in D8 was determined to go up to hundreds of keV.  The voltage was stepped down by 10 steps on Dec. 14 and again on Dec. 17 (days of year 349 and 352), leaving the voltage at 2617.6V.  D8 events remained off this week.
* 2012 December 19: The noise in D8 was determined to go up to hundreds of keV.  The voltage was stepped down by 10 steps on Dec. 14 and again on Dec. 17 (days of year 349 and 352), leaving the voltage at 2617.6V.  D8 events remained off this week.
-
* 2012 December 26: D8's voltage was stepped down another 10 steps on Dec. 21 (DOY 356).  This finally brought down the noise, which is now not present even coming out of the SAA.  The last easily noticeable noise in the rates was recorded on Dec. 19.  Detector 8 events will soon be turned back on.
+
* 2012 December 26: D8's voltage was stepped down another 10 steps on Dec. 21 (DOY 356) to 2475.5V.  This finally brought down the noise, which is now not present even coming out of the SAA.  The last easily noticeable noise in the rates was recorded on Dec. 19.  Detector 8 events will soon be turned back on.