- | + | * 15 Feb 2012: Final stages of the anneal, all detectors, are back to 4000 V, except for detector 9 (which showed some arcing after SAA passages). | |
+ | |||
+ | * 7 - 14 March 2012: Some noise was observed on detector 9. On March 8 at 01:24 UTC detector G9 HV was lowered by ten counts. Some sporadic noise was observed, but the detector appears to fixed it-self. | ||
+ | |||
+ | * 11 - 18 April 2012: | ||
+ | ** Rear fast threshold from 0x45 -> 0x60 | ||
+ | *** /IDPUDUMPTABL DIBTBL9 | ||
+ | *** /IDPUTABLE9 REARFASTDAC ;Load address 0x262C | ||
+ | *** /IDPULOAD VALUE=0x60 ;Byte 10, Rear Fast threshold |
Revision as of 06:27, 5 September 2012
- 15 Feb 2012: Final stages of the anneal, all detectors, are back to 4000 V, except for detector 9 (which showed some arcing after SAA passages).
- 7 - 14 March 2012: Some noise was observed on detector 9. On March 8 at 01:24 UTC detector G9 HV was lowered by ten counts. Some sporadic noise was observed, but the detector appears to fixed it-self.
- 11 - 18 April 2012:
- Rear fast threshold from 0x45 -> 0x60
- /IDPUDUMPTABL DIBTBL9
- /IDPUTABLE9 REARFASTDAC ;Load address 0x262C
- /IDPULOAD VALUE=0x60 ;Byte 10, Rear Fast threshold
- Rear fast threshold from 0x45 -> 0x60